Instruction Manual RX-I Fully automatic blood pressure monitor for measurement on the wrist
Gebrauchsanweisung RX-I Vollautomatisches Blutdruckmeßgerät zur Messung am Handgelenk
Gebruiksaanwijzing RX-I Volledig automatische bloeddrukmeter voor de pols
Manuel d’instructions RX-I Tensiomètre entièrement automatique pour prise de mesure au poignet
Manual de instrucciones RX-I Aparato completamente automático para medir la presión arterial en la muñeca
Manuale di instruzioni RX-I Misuratore della pressione per il polso completamente automatico
Vista de página 0
1 2 3 4 5 6 7 8 9

Resumo do Conteúdo

Página 1

1English Page 2 - 10Deutsch Seite 11 - 19Nederlands Pagina 20 - 28Français Page 29 - 37Italiano Pagina 38 - 46Español Página 47 - 55Instruction Man

Página 2 - Contents

English2The OMRON RX-I is a very compact, fully automatic wrist blood pressure monitor.Operating on the oscillometric principle, it measures your bloo

Página 3

English3Important:• RX-I Blood pressure measurements are not possible in cases of severearteriosclerosis (hardening of the arteries).• The pulse displ

Página 4 - 10 - 15mm

English42. Fitting the cuff10 - 15mmPlace the cuff on your left wrist with yourthumb pointing up. Align the edge of thecuff as shown in the diagram.Pl

Página 5

English5When you press the ON/OFF(O/I)button, allthe symbols on the display light up briefly.This serves to check the display. Then allthe symbols go

Página 6

English63. Memory functionThe measured and displayed blood pressure and pulse values are auto-matically stored in the memory.The memory can store the

Página 7 - Maintenance and storage

English7Possible faults, their causes and rectificationFaults and their possible causesIncorrect measurement isindicated by E , EE or E A corr

Página 8 - Technical data

English8CalibrationThe accuracy of this blood pressure monitor has been carefully tested and isdesigned for a long service life. It is generally recom

Página 9 - Organization

English9General information about blood pressureThe blood circulation is responsible for supplying the body with oxygen. Bloodpressure is the pressure